Dimensional accuracy and surface detail reproduction of two hydrophilic vinyl polysiloxane impression materials tested under dry, moist, and wet conditions

  • Cynthia S Petrie, DDS, MS

      Affiliations

    • Assistant Professor, Department of Restorative Dentistry, University of Missouri-Kansas City, School of Dentistry, Kansas City, Mo, USA
  • ,
  • Mary P Walker, DDS, PhD

      Affiliations

    • Associate Professor, Director of Dental Biomaterials, Department of Restorative Dentistry University of Missouri-Kansas City, School of Dentistry, Kansas City, Mo, USA
    • Corresponding Author InformationReprint requests to: Dr Mary P. Walker, University of Missouri-Kansas City, School of Dentistry, 650 E. 25th St, Kansas City, MO 64108, USA, Fax: (816) 235-5524
  • ,
  • Aisling M O'Mahony, BDS, DDS, MS

      Affiliations

    • Consultant in Restorative Dentistry, St James's Hospital, Dublin, Ireland
  • ,
  • Paulette Spencer, DDS, PhD

      Affiliations

    • Professor, Departments of Oral Biology and Pediatric Dentistry, Director of Graduate Studies and Research, Department of Oral Biology, University of Missouri-Kansas City, School of Dentistry, Kansas City, Mo, USA

Abstract 

Statement of problem

A major limitation of vinyl polysiloxane (VPS) impression materials is their hydrophobicity. There are 2 aspects to this problem, the wettability of the polymerized impression by dental gypsum materials and the ability of the unpolymerized material to wet intraoral tissues. To address this problem, manufacturers have added surfactants and labeled the new products as hydrophilic vinyl polysiloxane.

Purpose

The purpose of this investigation was to compare dimensional accuracy and surface detail reproduction of 2 hydrophilic VPS impression materials, when used under dry, moist, and wet conditions.

Material and methods

A total of 102 impressions were made of stainless steel metal dies similar to those described in American Dental Association (ADA) specification 19. The dies had 2 vertical and 3 horizontal lines inscribed on their superior surfaces. Impressions were made under dry, moist, and wet conditions. Dimensional accuracy was measured by comparing the average length of the middle horizontal line in each impression to the same line on the metal die, by use of a measuring microscope with an accuracy of 0.001 mm. A 2-way analysis of variance and least significant difference post hoc test were used to compare mean dimensional changes (α=.05). Surface detail reproduction was evaluated in 2 ways: (1) by use of criteria similar to ADA specification 19 for detail reproduction, continuous replication of at least 2 of the 3 horizontal lines, and (2) by use of a method developed for this study that categorized the impressions as satisfactory or unsatisfactory based on their surface characteristics: presence of pits, voids, or roughness. Pearson χ2 (α=.05) was used to compare detail reproduction results.

Results

Conditions (dry, moist, and wet) did not cause significant adverse effects on the dimensional accuracy of either material. The mean dimensional change and SD were 0.005% ± 0.002% or less. With both surface detail analyses, dry, moist, and wet conditions had a significant effect on the detail reproduction of both materials (P<.05). Only under dry conditions did both impression materials continuously replicate at least 2 of the 3 horizontal lines 100% of the time. Under moist conditions, 82% of the Aquasil impressions and 100% of the Reprosil impressions were judged satisfactory, while under wet conditions, only 47% Aquasil and 11% Reprosil impressions were satisfactory. With the additional surface detail characterization, only under dry conditions were impressions produced with clinically acceptable surface quality (Aquasil 77% and Reprosil 100% satisfactory).

Conclusions

Dimensional accuracy of both materials tested was well within ADA standards. Best surface detail results were obtained only under dry conditions for both materials.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 15.00 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Funded in part by the Rinehart Foundation, University of Missouri-Kansas City School of Dentistry.

PII: S0022-3913(03)00515-8

doi:10.1016/S0022-3913(03)00515-8

Journal of Prosthetic Dentistry
Volume 90, Issue 4 , Pages 365-372, October 2003